An ELISA was set up using polyvinylchloride microtiter plates coated with rabbit anti-UK IgG's and affino-purified goat anti-UK IgG's as second antibody. Detection occurred with rabbit anti-goat IgG antibodies conjugated with alkaline phosphatase. The assay is specific for urokinase (UK) with a detection limit of 100 pg/ml sample. Tissue-type plasminogen activator, up to concentrations of 100 ng/ml, does not interfere. The assay measures the antigen of the inactive zymogen pro-UK, the active enzyme UK and the UK-inhibitor complex with equal efficiency and gives the total UK antigen present, irrespective of its molecular form. Culture media of fibroblasts, endothelial- and kidney cells showed, despite the absence of active UK, antigen levels of 1.2, 23 and 65 ng/ml, respectively. In human plasma the UK concentration was found to be 3.5 +/- 1.4 ng/ml (mean +/- SD, n = 54). The inter- and intra-assay variations were 20% and 6%, respectively.
Organisatie | Hanze |
Gepubliceerd in | Thrombosis and Haemostasis Schattauer, Vol. 43, Uitgave: 5, Pagina's: 569-577 |
Datum | 1986-09-01 |
Type | Artikel |
DOI | 10.1016/0049-3848(86)90077-0 |
Taal | Engels |